Endress+Hauser CA80SI Silica Analyzer Liquiline System

SKU: 4695814-1-1035-817
In Stock

  • Primary Analysis: Silica Concentration Levels
  • Method: Colorimetric Heteropoly Blue
  • Sample Channels: Up to Six Channels
  • Power Supply: 100-240 VAC Universal

Did you like this product? Add to favorites now and follow the product.

Add to wishlist
Have a Question? Ask a Specialist

(+86) 132-6243-8118 Start Live Chat

Product Support

Description

Endress+Hauser CA80SI Silica Analyzer Liquiline System

The Endress+Hauser CA80SI silica analyzer is designed for continuous measurement of silica in aqueous solutions, especially in power generation and utility applications. By tracking trace silica in boiler feed water, steam, condensate, and other water-steam circuit points, it helps operators identify conditions that can contribute to glass-like deposits on turbines and other plant equipment. This early visibility supports timely process adjustments and helps protect equipment performance.

The analyzer is part of the Liquiline System and uses the standard heteropoly blue colorimetric method. Because this method is comparable with laboratory analysis, operators can more readily compare online readings with lab results during routine monitoring and validation.

Key Benefits of Silica Analysis

Early warning for plant protection

Silica monitoring helps assess boiler feed water purity and steam quality. It can also provide an early indication of condenser leaks, where silica-rich cooling water may enter the process, and help identify when ion exchangers are approaching exhaustion. These alerts give plant teams time to investigate and take appropriate corrective action.

In turn, timely monitoring can support more informed maintenance planning and process decisions, including the optimization of blowdown and ion-exchange regeneration schedules. The Endress+Hauser CA80SI silica analyzer helps users respond to changing silica levels before they become a larger operational concern.

Measurement method and dependable sample handling

The CA80SI uses colorimetric measurement based on the heteropoly blue method. Its combination of peristaltic and high-precision dispenser pumps supports stable sample delivery and reagent dosing, helping the analyzer provide consistent readings while keeping maintenance requirements low.

Automatic calibration and cleaning are available, and users can configure measuring, cleaning, and calibration intervals to suit their process needs. The analyzer is intended for particle-free samples, with a sample flow rate of 60 to 250 ml/min.

Flexible Configuration and System Integration

The analyzer offers two measuring ranges: 0 to 500 µg/l (ppb) and 50 to 5000 µg/l (ppb). Depending on the configuration, it can support one, two, four, or six measuring channels, allowing users to monitor several sample points with one system.

For changing monitoring requirements, the Endress+Hauser CA80SI silica analyzer can be expanded with up to six sample channels. It can also connect to as many as four digital Memosens sensors, providing an upgrade path toward a broader measuring station. Available communication options include two 0/4 to 20 mA outputs, with optional Webserver, Modbus, EtherNet/IP, and PROFIBUS DP connectivity for integration into process control systems.

Advanced diagnostics and remote access via a web server can help teams investigate errors more efficiently. The optional Liquiline Edge Module CYY7 can transfer process data to Netilion independently of the fieldbus, enabling remote data overview and access to IIoT services.

Typical Applications in Power and Utilities

The CA80SI is intended for silica determination in boiler feed water and for monitoring water-steam circuits and condensate. It can also support monitoring and optimization in reverse osmosis and desalination processes. In power plants and utilities, these measurements help maintain water quality and identify changes that may affect turbines, ion exchangers, and related equipment.

Its process conditions include a temperature range of 5 to 45 °C and a pressure range of 1 to 5 bar. The ambient temperature range is 5 to 40 °C. Confirm that the sample conditions, selected measuring range, and channel configuration match the intended installation before specifying the analyzer.

Why Buy from TPT24

TPT Instrumentation Co., Ltd. has served instrumentation customers since 2017 through tpt24.com. With its headquarters in Shenzhen and regional offices in Munich and Dubai, TPT24 supports customers in more than 60 countries with instrumentation sourcing and technical consultation.

When considering the Endress+Hauser CA80SI silica analyzer, buyers can contact TPT24 for product information and quotation support. TPT24’s stated business principles are honesty, fairness, and service, with a focus on helping customers identify suitable measurement equipment for their applications.

Ordering Support for the CA80SI

The Endress+Hauser CA80SI silica analyzer combines heteropoly blue colorimetric analysis, configurable sample channels, and optional digital connectivity for demanding water-quality monitoring tasks. Review the measurement ranges, sample requirements, process conditions, and integration options against your plant needs. Contact the technical team at tpt24 to discuss the analyzer and request a quotation.

Specifications

SpecificationDetails
Measuring principleColorimetric
CharacteristicAnalyzer for silica in aqueous solutions
Measuring methodComply with standard colorimetric measuring principle – heteropoly blue method
SizeHousing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in
DesignOpen design / cabinet
Plastic ASA-PC
Process temperature5 to 45 °C (41 to 113 °F)
Ambient temperature5 to 40 °C (41 to 104 °F)
Process pressure1 to 5 bar (14.5 to 72.5 psi)
Sample flow rate60 to 250 ml/min (2.03 to 8.45 fl.oz/min)
Consistency of the sampleParticle free
SpecialsEasy upgrade to measuring station with up to four digital Memosens sensors
Automatic calibration and cleaning
User-configurable measuring, cleaning and calibration intervals
optional up to 6 sample channels
Flexible functionality and modular expandable
Digital communication for remote access
ApplicationDetermination of silica concentration in boiler feed water
Monitoring of water-steam-circuit and condensate
Monitoring and optimization of reverse osmosis and desalination
Power supply100 to 120 VAC / 200 to 240 VAC ± 10%
50 ± 1 or 60 ± 1.2 Hz
Output / communication2x 0/4 to 20 mA
Optional: Webserver, Modbus, Ethernet/IP, Profibus DP
Input1, 2, 4 or 6 measuring channels
Optional 1 to 4 digital inputs for sensors with Memosens protocol
Measuring range0 to 500 µg/l (ppb)
50 to 5000 µg/l (ppb)
ConsumablesReagents and standard solutions CY80SI
Maintenance kit CAV800

Data Sheet

Download Data Sheet Of:Endress+hauser CA80SI Silica analyzer Liquiline System