Endress+hauser CLS16E Digital conductivity sensor Memosens

SKU: 4695814-1-1035-539
In Stock

  • Certified hygienic sensor design
  • Memosens 2.0 IIoT ready
  • Sterilizable and autoclavable
  • High precision measurement

Did you like this product? Add to favorites now and follow the product.

Add to wishlist
Have a Question? Ask a Specialist

(+86) 132-6243-8118 Start Live Chat

Low Prices

Price match guarantee

Guaranteed Fitment.

Always the correct part

In-House Experts.

We know our products

Easy Returns.

Quick & Hassle Free

Description

Endress+hauser CLS16E Digital conductivity sensor Memosens

The Endress+Hauser Memosens CLS16E represents the pinnacle of hygienic conductivity sensing, designed for applications where the utmost precision and purity are non-negotiable. This high-end sensor boasts a certified hygienic design, rigorously conforming to FDA and USP Class VI standards, ensuring its suitability for the most sensitive processes within the life sciences and food industries. It delivers measurements with exceptional precision, providing reliable data essential for maintaining optimal process conditions and guaranteeing superior product quality. The CLS16E is built for durability, offering a long operating life, partly due to its innovative replaceable seal design. Furthermore, its integration with Memosens 2.0 technology allows for the storage of extensive calibration and process data, which is instrumental in enabling predictive maintenance strategies and forming the robust foundation required for advanced IIoT services, making it a truly future-ready solution.

Hygienic Applications In High Purity Systems

The Memosens CLS16E sensor is specifically engineered for critical applications within pure and ultrapure water systems, where meticulous monitoring and control are paramount. Its primary function is in the production and monitoring of Water for Injection (WFI), a vital component in pharmaceutical manufacturing, ensuring the highest standards of sterility and purity. It plays a crucial role in Electro-deionization (EDI) systems, efficiently monitoring water conductivity to optimize the deionization process and maintain water purity. In ion exchange systems, the CLS16E accurately tracks the performance of resins, signaling when regeneration is needed to maintain ion removal efficiency. It is also extensively used in reverse osmosis (RO) plants to assess membrane performance and confirm the production of high-purity water streams. Additionally, the sensor is applied in distillation processes, verifying the purity of the final distilled product. Its robust design and hygienic certifications make it ideal for these demanding applications, ensuring process integrity and product safety.

Certified Hygienic Design For Sensitive Processes

The Memosens CLS16E sensor is distinguished by its exceptional adherence to the strictest hygienic standards. It is certified with a comprehensive quality certificate and EN 10204 3.1, further solidifying its credentials. Critically, it complies with leading industry standards such as EHEDG (European Hygienic Engineering & Design Group), USP Class VI (United States Pharmacopeia), ASME (American Society of Mechanical Engineers), and FDA (Food and Drug Administration). This extensive list of certifications underscores its suitability for the most demanding hygienic applications, particularly in the life sciences and food industries. These certifications guarantee that the sensor’s design, materials, and performance meet stringent requirements for preventing microbial contamination and ensuring product safety, making it a trusted component in critical pharmaceutical, biopharmaceutical, and food processing environments where cleanliness and sterility are paramount.

Sterilizable And Autoclavable For Ultimate Cleanliness

Ensuring the highest levels of hygiene and preventing contamination are critical in many industries, and the Memosens CLS16E is designed to meet these exacting demands. The sensor is not only sterilizable and autoclavable but also highly resistant to Cleaning In Place (CIP) and Sterilization In Place (SIP) procedures. This means it can withstand rigorous cleaning and sterilization cycles at high temperatures and pressures without compromising its performance or integrity. Its construction, featuring robust stainless steel, ensures excellent corrosion resistance, contributing to its durability even under harsh cleaning conditions. The ability to undergo sterilization and CIP/SIP processes makes the CLS16E an indispensable tool for applications in the pharmaceutical, biotechnology, and food and beverage sectors where maintaining a sterile environment is essential for product safety and regulatory compliance, providing peace of mind and operational reliability.

IIoT Ready With Memosens 2.0 Advanced Data

The Memosens CLS16E is at the forefront of smart sensor technology, fully embracing the Industrial Internet of Things (IIoT) with its integrated Memosens 2.0 capability. This advanced digital platform allows the sensor to store a significantly larger volume of calibration and process data compared to traditional sensors. This enhanced data logging facilitates predictive maintenance, enabling users to analyze trends, anticipate potential sensor drift or failure, and schedule maintenance proactively, thereby minimizing costly unscheduled downtime. The detailed historical data provides a comprehensive overview of the sensor’s performance over time. This rich dataset forms the perfect basis for sophisticated IIoT services, allowing for remote monitoring, advanced analytics, and seamless integration with plant-wide digital systems, ensuring that operations are not only efficient but also intelligent and future-proofed against evolving industrial demands.

Precision Measurement Under Demanding Conditions

Regardless of the operating environment, the Memosens CLS16E consistently delivers highest precision and measuring accuracy. This exceptional performance is maintained even when subjected to high temperatures and pressures, ensuring that the data you receive is always reliable for optimal process and product quality. The sensor’s design minimizes potential interference, and its conductive measuring principle, utilizing a high-quality, electropolished stainless steel electrode, guarantees accurate readings across its specified range. The robust, corrosion-free construction from stainless steel further contributes to its longevity and consistent performance. This unwavering accuracy under challenging conditions is crucial for industries where even minor deviations can have significant consequences, providing users with the confidence that their processes are under precise control at all times.

Specifications

  • Measuring principle

    Conductive

  • Application

    Pure and ultrapure water
    Pharmaceutical industry
    Final rinse
    WFI

  • Characteristic

    Hygienic 2-electrode conductivity sensor

  • Measuring range

    k=0,1: 0,04 to 500 µS/cm

  • Measuring principle

    Conductive conductivity cell with electropolished stainless steel electrodes

  • Design

    Hygienic 2-electrode conductivity sensor with coaxially arranged electrodes, electropolished

  • Material

    n contact with medium
    Sensor depending on order execution:
    Electropolished, stainless steel 1.4435 (AISI 316L)
    PEEK
    Sealing depending on order execution:
    Form seal ISOLAST (FFKM)
    Form seal EPDM

  • Dimension

    Electrode diameter: 17 mm (0.66 inch)
    Electrode length: 54 mm (2.13 inch)

  • Process temperature

    -5 to 120 °C (23 to 248 °F)
    For Sterilizsation: max. 150 °C at 5 bar
    (Max. 302 °F at 73 psi)

  • Process pressure

    13 bar at 20 °C (188 psi at 68 °F)
    9 bar at 120 °C (130 psi at 248 °F)

  • Temperature sensor

    Pt1000

  • Ex certification

    ATEX, NEPSI, CSA, IECEx, INMETRO, EAC Ex

  • Connection

    Process connection: Varivent, Neumo, BioControl, Clamp
    Sensor connection: Inductive, digital connection head with Memosens 2.0 technology

  • Ingress protection

    IP68

  • Additional certifications

    Material certification 3.1
    EHEDG certified, hygienic design
    CoC ASME BPE
    Conformity to cGMP derived

Data Sheet

Download Data Sheet Of:Endress+hauser CLS16E Digital conductivity sensor Memosens