Endress+hauser CPS96E Digital combined pH/ORP sensor Memosens
- Measures: pH & ORP
- Aperture: Open, Fouling Resistant
- Reference: Ion Trap, Gel
- Process Temp: 0 to 110°C
Did you like this product? Add to favorites now and follow the product.
(+86) 132-6243-8118 Start Live Chat
Description
Endress+Hauser CPS96E: Tough pH/ORP for Solids
Specialist for Media with High Suspended Solids
The Memosens CPS96E is meticulously engineered as a specialist for the most challenging media, particularly those characterized by a high content of suspended solids. This includes complex mixtures like dispersions, precipitations, and emulsions, which often pose significant measurement difficulties for conventional sensors. Its innovative open aperture design is a key feature, proactively preventing fouling and ensuring reliable, simultaneous measurement of both pH and ORP. This robust capability makes it an indispensable tool for maintaining process stability and quality in highly contaminated environments.
Memosens 2.0 Enhances Process Integrity and Data
Integrating advanced Memosens 2.0 digital technology, the CPS96E achieves a powerful combination of maximum process integrity and simplified operation. This state-of-the-art digital interface ensures that the sensor is highly resistant to moisture, enabling convenient and precise calibration in laboratory settings rather than in situ. Crucially, Memosens 2.0 provides extended storage capabilities for both calibration and critical process data. This comprehensive data logging forms the perfect foundation for implementing effective predictive maintenance strategies and leveraging enhanced IIoT services.
Reliable pH/ORP in Chemical and Paper Industries
The CPS96E is ideally suited for a broad range of demanding industrial applications, particularly within chemical processes and the pulp and paper industry. In these sectors, media often contain significant levels of suspended solids, making accurate and stable pH and ORP measurements challenging. Its robust design and open aperture system are specifically developed to overcome these hurdles, ensuring consistent performance in highly contaminated media such as slurries, pigment dispersions, and wastewater, which are common in these heavy industries.
Benefits: Stable Measurements in Contaminated Media
A key benefit of the CPS96E is its ability to deliver stable and accurate measurements even in heavily contaminated media with high concentrations of solids, emulsions, or precipitates. The unique design, featuring an open aperture and a double gel reference with an ion trap, provides exceptional resistance to poisoning. This robust construction, combined with advanced gel technology, ensures that the sensor maintains its performance integrity over extended periods, minimizing measurement drift and extending the sensor’s operational lifespan in adverse conditions.
Simultaneous pH, ORP, and Reference Impedance Monitoring
The CPS96E offers the distinct advantage of simultaneous measurement of pH, ORP, and rH values (in rH mode), providing a holistic view of the process chemistry for tighter control. Beyond these critical parameters, an additional platinum electrode allows for constant monitoring of the reference impedance. This continuous monitoring is a powerful diagnostic tool, enabling users to assess the sensor’s quality and condition in real-time. This capability ensures proactive maintenance, enhancing reliability and reducing unexpected downtime.
Robust Design Unaffected by Pressure and Temperature Swings
The innovative design of the CPS96E ensures that its performance remains stable and unaffected by fluctuations in process pressure and temperature. The open aperture is crucial in this regard, as it actively prevents the build-up of diffusion potential at the junction, a common issue that can compromise accuracy in conventional sensors under variable conditions. This robust engineering guarantees reliable and consistent measurement results, making the CPS96E a highly dependable choice for dynamic and challenging process environments where stability is paramount.
Certified Safe for Hazardous Area Operations Globally
Safety is a critical factor in many industrial settings, and the Endress+Hauser Memosens CPS96E proudly holds comprehensive international hazardous area certifications. These include ATEX, IECEx, CSA C/US, NEPSI, Japan Ex, and INMETRO approvals. Such extensive certification permits its safe and reliable use in hazardous zones 0, 1, and 2, which encompass environments with potentially flammable gases, vapors, or combustible dust. This broad compliance ensures that the CPS96E can be deployed globally in high-risk industrial applications with confidence.
Your Premier Global Source: tpt24 for CPS96E
For industries worldwide seeking the high-performance Endress+Hauser Memosens CPS96E digital pH/ORP sensor, tpt24 is the undisputed premier global online retailer. We are deeply committed to ensuring that every customer receives an unparalleled guarantee of authenticity and superior quality with their CPS96E purchase. tpt24 provides secure and efficient worldwide delivery, meaning no matter your geographical location, you can confidently procure this essential instrumentation. Trust tpt24 for genuine products and exceptional service to meet your most demanding process needs.
Specifications
-
Measuring principle
Potentiometric
-
Application
pH / ORP sensor for heavily contaminated media
and suspended solids
• Chemical processes
• Pulp and paper industry
• Flue gas cleaning
• Contaminated media:
• Solids
• Emulsions
• Precipitation reactions -
Characteristic
Gel pH/ORP electrode with open aperture and double gel reference with ion trap for chemical applications
-
Measuring range
ORP: –1 500 to 1 500 mV
pH: 0 to 14 -
Measuring principle
Pt-disc as additional ORP element
rH measurement and control of the reference impedance -
Design
All shaft lengths with temperature sensor
Advanced gel technology
Hydrolysis stable gel -
Material
Sensor shaft: Glass to suit process
pH membrane glass: Type B
Metal lead: Ag/AgCl
ORP measuring element: Platinum
O-ring: FKM
Process coupling: PPS fiber-glass reinforced
Nameplate: Ceramic metal oxide -
Dimension
Diameter: 12 mm (0.46 inch)
Shaft lengths: 120, 225, 360 and 425 mm
(4.68, 8.77, 14.04 and 16.57 inch) -
Process temperature
0 to 110 °C (32 to 230 °F)
-
Process pressure
0.8 to 14 bar (11.6 to 203 psi) absolute
-
Temperature sensor
NTC 30k
-
Ex certification
With ATEX, IECEx, CSA C/US, NEPSI, Japan Ex and INMETRO approvals for use in
hazardous areas Zone 0, Zone 1 and Zone 2 -
Connection
Inductive, digital connection head with Memosens 2.0 technology
-
Ingress protection
IP68