Endress+Hauser Promass Q 500 Proline Promass Coriolis Flowmeter

SKU: 4695814-1-1035-881
In Stock

  • Mass flow error ±0.05%
  • Remote transmitter with 4 I/Os
  • Optimized for entrained gas
  • Wide operating temperatures

Did you like this product? Add to favorites now and follow the product.

Add to wishlist
Have a Question? Ask a Specialist

(+86) 132-6243-8118 Start Live Chat

Product Support

Description

Endress+hauser Q 500 Proline Promass Coriolis flowmeter

Industrial instrumentation requires uncompromising precision, long-term reliability, and advanced diagnostic capabilities across complex fluid processes. The Promass Q 500 stands as the pinnacle of Coriolis measurement technology, engineered to deliver unsurpassed accuracy for mass flow, volume flow, and density. As the ultimate choice for critical custody transfer applications where legally verifiable transactions are paramount, the advanced Promass Q 500 ensures operational fairness and minimizes financial discrepancies. Facility operators rely on the Promass Q 500 to maintain strict process control under extreme operating conditions.

Coriolis Principle and Fluid Independence

At the core of this device is the Coriolis measuring principle, which operates with remarkable independence from fluctuating fluid properties such as viscosity, density, and temperature. Unlike traditional flowmeters influenced by changing process conditions, the Promass Q 500 maintains consistent metrological stability. This fundamental design allows the Promass Q 500 to deliver precise mass and volume flow measurements with an outstanding error rate of plus or minus 0.05 percent when using PremiumCal.

Advanced MFT for Entrained Gas Conditions

Challenging multiphase processes often suffer from measurement instability caused by gas bubbles. The Promass Q 500 effectively overcomes this challenge through proprietary Multi-Frequency Technology (MFT). This specialized feature stabilizes readings in aerated fluids, ensuring continuous process reliability.

Promass Q 500

Smart Diagnostics with Heartbeat Technology

Plant efficiency depends on proactive maintenance and verified instrument health. The Promass Q 500 integrates innovative Heartbeat Technology for continuous self-monitoring and verification without process interruption.

Proline 500 Transmitter Connectivity

The Proline 500 remote transmitter provides flexible mounting options and extensive fieldbus compatibility, including HART, PROFIBUS, PROFINET, Modbus, and EtherNet/IP, alongside integrated WLAN access and touch control.

Simultaneous Multivariable Measurement Capabilities

Beyond high-precision mass flow and volume flow measurement, the versatile Promass Q 500 provides simultaneous monitoring of fluid density and process temperature. By capturing these critical parameters within a single physical measurement point, plant engineers eliminate the need for multiple discrete instruments, significantly reducing installation complexity and capital expenditure. The high-performance signal processing unit calculates corrected volume and concentration variables on the fly, making the Promass Q 500 an exceptional asset for complex chemical processing, oil and gas distribution, and food and beverage production lines where fluid composition changes dynamically.

Robust Mechanical Design and Pipeline Integration

Industrial environments subject field instruments to severe thermal shocks, external pipe loads, and mechanical vibrations. The rugged sensor housing of the Promass Q 500 is engineered with balanced dual-tube measuring systems and heavy-duty process connections that completely isolate internal measurements from external pipeline stresses. Furthermore, the self-draining sensor geometry ensures complete cleanability in sanitary installations and prevents product accumulation or contamination during CIP and SIP cleaning cycles. Operators can install the Promass Q 500 in tight spaces or vertical lines without compromising metrological performance, ensuring maximum flexibility during plant revamping projects.

Hazardous Area Approvals and Global Standards Compliance

Safety and regulatory compliance remain paramount in process industries handling flammable media or operating within explosive atmospheres. The Promass Q 500 holds extensive international hazardous area certifications, including ATEX, IECEx, and cCSAus, allowing safe deployment in Zone 1 and Zone 2 hazardous locations. Additionally, rigorous calibration traceable to international standards guarantees absolute legal compliance for fiscal metering and custody transfer operations across global jurisdictions.

Reliable Global Instrumentation Supply by TPT24

Heavy-duty process equipment demands guaranteed hardware integrity from dispatch to field installation. Operating since 2017 under TONGLENG Precision Tools, TPT Instrumentation Co., Ltd. has developed specialized export channels serving industrial operators across sixty countries worldwide. TPT24 acts as your direct sourcing partner for genuine Endress+Hauser devices, ensuring strict packaging protocols, factory documentation tracking, and reliable logistics management. From initial specification review to international port-to-door delivery, we simplify industrial procurement while protecting mission-critical delivery milestones.

Specifications

SpecificationDetails
Measuring principleCoriolis
Product headlineInnovative specialist for challenging applications, as remote version with up to 4 I/Os.
Secured measuring quality – unmatched accuracy of mass flow, volume flow and density.
Highest measurement performance for custody transfer, density and under tough process conditions.
Sensor featuresOptimized performance for liquids with entrained gas – MFT (Multi-Frequency Technology). Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs.
Mass flow: measurement error ±0.05 % (PremiumCal). Density: measurement error ±0.2 kg/m³. High turndown due to low pressure loss/zero point.
Transmitter featuresFull access to process and diagnostic information – numerous, freely combinable I/Os and fieldbuses. Reduced complexity and variety – freely configurable I/O functionality. Integrated verification – Heartbeat Technology.
Remote version with up to 4 I/Os. Backlit display with touch control and WLAN access. Standard cable between sensor and transmitter.
Nominal diameter rangeDN 25 to 250(1 to 10″)
Wetted materialsMeasuring tube: 1.4404 (316/316L); stainless steel for cryogenic applications
Connection: 1.4404 (316/316L)
Measured variablesMass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
Max. measurement errorMass flow (liquid): ±0.10 % (standard), 0.05 % (option)
Volume flow (liquid): ±0.10 %
Mass flow (gas): ±0.25 %
Density (liquid): +/-0.2 kg/m³ (standard), +/-0.1 kg/m³ (option)
Measuring range0 to 2 400 000 kg/h (0 to 88183 lb/min)
Max. process pressurePN 100, Class 600, 63K
Medium temperature rangeStandard: –50 to +205°C (–58 to +401 °F)
Option: –196 to +150 °C (–321 to +302 °F)
Ambient temperature rangeStandard: – 4 to +60 °C (–4 to +140 °F)
Option:–60 to +60 °C (–76 to +140 °F)
Sensor housing material1.4404 (316L), highest corrosion resistance
Sensor connection housing (standard): AlSi10Mg, coated
Sensor connection housing (option): 1.4301 (304); 1.4404 (316L); 1.4409 (CF3M) similar to 316L
Transmitter housing materialAlSi10Mg, coated; 1.4409 (CF3M) similar to 316L; Polycarbonat
Degree of protectionSensor remote version (standard): IP66/67, type 4X enclosure
Sensor remote version (option): IP69. Transmitter remote version: IP66/67, Type 4X enclosure
Display/Operation4-line backlit display with touch control (operation from outside)
Configuration via local display and operating tools possible
Outputs4 outputs:
4-20 mA HART (active/passive)
4-20 mA WirelessHART
4-20 mA (active/passive)
Pulse/frequency/switch output (active/passive)
Double pulse output (active/passive)
Relay output
InputsStatus input
4-20 mA input
Digital communicationHART, PROFIBUS DP, PROFIBUS PA, FOUNDATION Fieldbus, Modbus RS485, PROFINET, PROFINET over Ethernet-APL, Ethernet/IP, OPC-UA
Power supplyDC 24 V
AC 100 to 230 V
AC 100 to 230 V / DC 24 V (non-hazardous area)
Hazardous area approvalsATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, KC
Product safetyCE, C-tick, EAC marking
Functional safetyFunctional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
Metrological approvals and certificatesCalibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
MI-005 Liquids other than water (Hydrocarbons, Liquified gases, Cryogenic liquids)
OIML R117 (Liquids other than water, Liquified gases, Cryogenic liquids)
NTEP (Liquids other than water, Cryogenic liquids)
MC (Liquids other than water, Cryogenic liquids)
Marine approvals and certificatesLR approval, DNV approval, ABS approval, BV approval, CCS approval
Pressure approvals and certificatesPED, CRN, AD 2000
Material certificates3.1 material
NACE MR0175/MR0103, PMI; welding test acc. to EN ISO, ASME, NORSOK
Hygienic approvals and certificates3-A, EHEDG, cGMP

Data Sheet

Download Data Sheet Of:Endress+hauser Q 500 Proline Promass Coriolis flowmeter