Hanna HI932 Advanced Automatic Potentiometric Titrator

In Stock

  • pH range: −2.000 to 20.000 pH
  • Burette sizes: 5 to 50 mL
  • Dosing pump: 40,000‑step precision drive
  • Display: 5.7” color LCD

Did you like this product? Add to favorites now and follow the product.

Add to wishlist
Have a Question? Ask a Specialist

(+86) 132-6243-8118 Start Live Chat

Low Prices

Price match guarantee

Guaranteed Fitment.

Always the correct part

In-House Experts.

We know our products

Easy Returns.

Quick & Hassle Free

Description

Elevating Laboratory Precision with HI932 Innovation

The Hanna HI932 Advanced Automatic Potentiometric Titrator is a state‑of‑the‑art analytical instrument designed to take titration testing to an entirely new level. Built to surpass industry requirements, its cutting‑edge technology enables precise, repeatable, and customizable measurement for pH, mV, and ion‑selective electrode (ISE) analysis. Encased in a durable Cycoloy® body and utilizing chemically resistant PTFE tubing, the HI932 stands resilient against harsh laboratory chemicals and demanding workflows. Compact in design, it occupies 50% less bench space compared to earlier models, making it ideal for laboratories with limited real estate. Equipped with intuitive menus, high‑contrast backlit LCD display, and advanced dosing mechanics, the HI932 blends accuracy, durability, and user‑friendly operation into one cohesive, reliable instrument.

Dynamic Precision with Forty‑Thousand Step Dosing

At the core of HI932’s exceptional performance is its unmatched 40,000‑step piston‑driven dosing pump. This pump delivers extremely small and highly accurate titrant volumes, ensuring exact endpoint detection whether using fixed values or equivalence point determination by derivative curves. Flow rates range from 0.3 mL to twice the burette volume per minute, allowing flexible operation across different titration methods. The titrator supports burette sizes from 5 mL to 50 mL, with resolutions as fine as 0.001 mL and dosing accuracy up to ±0.005 mL on smaller burettes. Automatic burette size recognition streamlines setup for quick switching between methods, while PTFE chemical‑resistant aspiration and dispensing lines maintain reagent integrity. This precise hardware ensures reproducibility even with complex titration sequences, vital for analytical validity.

Versatile Measurement Across Multiple Analytical Modes

The HI932 is more than just a titrator—it is a multi‑function analytical station. It acts simultaneously as a pH meter, mV/ORP meter, and ISE meter, allowing complete electrochemical profiling on a single sample. Analytical flexibility covers acid/base titrations (pH or mV modes), redox, precipitation, complexometric, non‑aqueous, ion‑selective, and argentometric titration types. The endpoint detection can be set to fixed mV or pH values or determined via first or second derivative equivalence point calculation. Direct measurement modes include precision pH, ORP, and ion‑specific concentration readings. Calibration routines support up to five points with standard or custom buffers or standards, ensuring confidence in results for any analyte. This versatility is indispensable for laboratories serving multiple industries.

Exclusive Applications Serving Wide Range of Industries

Thanks to its adaptable design, the HI932 finds application in diverse fields requiring precision chemical analysis. In food and beverage manufacturing, it ensures quality control over acidity, salt content, and ingredient stability. Pharmaceutical laboratories use it for rigorous assay validation of active compounds and excipient properties. Environmental monitoring agencies deploy it for water quality analysis, detecting pollutants or nutrient concentrations with high reliability. Metallurgical and plating operations rely on it for precise measurement of bath compositions and to prevent defects in coatings. Educational and research institutions benefit from its repeatable methodology for teaching quantitative chemical analysis. When paired with the HI922 Autosampler, its reach expands even further, enabling correct sequential titration of up to 18 samples automatically, saving time and maximizing throughput.

Advanced Data Management and GLP Traceability Features

The HI932 ensures full compliance with Good Laboratory Practice (GLP) requirements by automatically storing comprehensive metadata for every titration. Reports capture sample identification, operator names, company, date, time, electrode ID codes, and calibration data. Up to 100 methods (user‑defined or standard) and 100 titration and pH/mV/ISE reports can be stored internally, alongside 30 autosampler sequences or 40 tray reports. Real‑time and stored graphs include mV‑volume or pH‑volume titration curves and derivative plots. USB host and PC connectivity allow export of methods and results in LIMS‑compatible formats. Direct printer connection enables hard copy generation, while analytical balance interfacing provides automated sample weight entry—minimizing manual error. These data‑management capabilities secure analytical integrity and audit readiness.

Ergonomic Design Enhancing Laboratory Workflow Efficiency

A key advantage of the HI932 is its focus on operator convenience. The 5.7‑inch color LCD with adjustable backlight offers wide viewing angles, ensuring readability from any bench position. The electrode holder accommodates up to three electrodes, four reagent tubes, and a temperature probe, mounted directly to the titrator body for stability. An overhead propeller stirrer with adjustable speeds from 200 to 2500 RPM guarantees sample homogeneity throughout measurement. Spill management is integrated via exterior gutters and interior trays protecting electronics. The titrator’s small footprint maximizes laboratory workspace, and intuitive on‑screen navigation with virtual keys eliminates training complexity. Users can perform tasks concurrently while titrations run, boosting productivity and freeing time for other critical analyses.

Seamless Integration with HI922 Autosampler Accessory

For laboratories requiring sample batch processing, pairing the HI932 with the HI922 Autosampler unlocks automation at its finest. Together, they can perform titrations in sequence for multiple samples, run up to eighteen automated analyses consecutively, and manage reagent addition via up to three peristaltic pumps. The autosampler integrates directly with the titrator, with tray and beaker identification to prevent mix‑ups. Status indicator lights keep operators informed of progress or alert them to issues, improving efficiency and reducing downtime. This combination is ideal for high‑throughput labs or for quality control departments that require consistent, traceable analysis without constantly monitoring instruments during operation.

Global Availability Through tpt24.com Excellence

When sourcing the Hanna HI932 Advanced Automatic Potentiometric Titrator, tpt24.com stands out as the premier global online store. It delivers genuine Hanna Instruments products with full authenticity and quality guarantees to any country worldwide. Customers benefit from secure international shipping, expert assistance, and warranty coverage, ensuring peace of mind with every purchase. Whether upgrading from manual titration or expanding laboratory capacity, professionals can trust tpt24.com to supply the HI932 ready for rigorous demands. Combined with Hanna’s engineering excellence, this makes for a solution that is both technically superior and logistically dependable—an asset for any laboratory, manufacturing site, or research facility seeking reliable titration control.

Specifications

pH Range −2.0 to 20.0 / −2.00 to 20.00 / −2.000 to 20.000 pH
pH Resolution 0.1 / 0.01 / 0.001 pH
pH Accuracy (@25°C) ±0.001 pH
pH Calibration Up to five points, standard/custom buffers
mV Range −2000.0 to 2000.0 mV
mV Resolution 0.1 mV
mV Accuracy (@25°C) ±0.1 mV
mV Calibration Single‑point offset
ISE Range 1×10⁻⁶ to 9.99×10¹⁰
ISE Resolution 1 / 0.1 / 0.01
ISE Accuracy (@25°C) ±0.001 pH
ISE Calibration Up to five points
Temperature Range −5.0 to 105.0°C / 23.0 to 221.0°F / 268.2 to 378.2 K
Temperature Resolution 0.1°C / 0.1°F / 0.1K
Temperature Accuracy (@25°C) ±0.1°C / ±0.2°F / ±0.1K
Stirrer Type and Speed Overhead propeller, 200 to 2500 RPM
Burette Sizes 5 / 10 / 25 / 50 mL
Burette Resolution 1/40000
Dosing Accuracy ±0.005 mL (5 mL), ±0.010 mL (10 mL), ±0.025 mL (25 mL), ±0.050 mL (50 mL)
Flow Rate 0.3 mL to 2 × burette volume/min
Endpoint Determination Fixed pH/mV or equivalence point (1st/2nd derivative)
Methods Capacity Up to 100 standard/user methods
Data Reports Up to 100 titration & pH/mV/ISE reports
Display 5.7” (320×240 pixel) backlit color LCD
Dimensions 315 × 205 × 375 mm
Weight Approx. 4.3 kg
Languages English, Portuguese, Spanish, Deutsch
Power Supply 100–240 VAC, US (type A) or EU (type C) plugs
Warranty 2 years